Primary reference in terms of air kerma for ionizing radiation fields produced by an electrostatic electron accelerator
1
CEA, LIST, Laboratoire National Henri Becquerel (LNE-LNHB), CEA-Saclay, 91191 Gif sur Yvette Cedex, France
2
ATRON METROLOGY, 14 allée des vindits, parc d'activité des fourches, 50130 Cherbourg en Cotentin, France
3
CEA, LIST, DM2I SCI LCAE, CEA-Saclay, 91191 Gif sur Yvette Cedex, France
Based on a radiation field produced by an electrostatic accelerator, a radiation survey meter test and calibration facility has been designed and characterized in terms of air kerma and ambient dose equivalent. The electron beam impinges a tantalum target to produce X-rays. The spectrum has been measured and calculated. Traceability to the International System of units is achieved by means of a calibration with a primary dosimeter for air kerma.