Calibration of primary measuring converter by using of modeling and experimental evaluation of the initial data
1 National Scientific Centre „Institute of Metrology“, 42, Myronosytska str., 61002, Kharkiv, Ukraine
2 O.M. Beketov National University of Urban Economy in Kharkiv, 17, Marshal Bazhanov str., 610002, Kharkiv, Ukraine
* Corresponding author: vladimir.skliarov@metrology.kharkov.ua
The main target of this work is to decrease the uncertainty of measurement of laser beam power by isoperibolic calorimeters. Measurement and calibration operating regimes of the isoperibolic calorimeter are investigated experimentally. Temperature transients during the heating and cooling of the calorimeter core are studied. The differences of transient parameters in measurement and calibration operating regimes of the calorimeter are described. Initial data for design of the calibration model of the isoperibolic calorimeter are determined. It is recommended to take into account the nonlinear thermal and physical characteristics of the core materials.
© The Authors, published by EDP Sciences, 2017
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 4.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.