Numéro |
2013
16th International Congress of Metrology
|
|
---|---|---|
Numéro d'article | 18001 | |
Nombre de pages | 5 | |
Section | Rayonnement optique / Optical radiation | |
DOI | https://doi.org/10.1051/metrology/201318001 | |
Publié en ligne | 7 octobre 2013 |
EMRP-ENV03: Traceability for surface spectral solar ultraviolet radiation
1 LCM (LNE-CNAM), 29 avenue Roger Hennequin, ZA Trappes-Elancourt, 78197 Trappes, France
2 LNE, 29 avenue Roger Hennequin, ZA Trappes-Elancourt, 78197 Trappes, France
a Corresponding author: jimmy.dubard@lne.fr
The ENV03 Solar UV project is aimed at improving the traceability of solar UV spectral irradiance measurement. This is achieved by developing new characterization methods and devices (stray light improved array spectroradiometer, Fourier Transform Spectrometer). The workpackages of the project are presented. We describe the output of two tasks in which the LNE is involved: the realization of a software for evaluating the measurement uncertainty using Monte Carlo technique, and the development of a reduced stray light array spectroradiometer fitted with band-pass filters to measure the UV spectral irradiance with an uncertainty <2%.
© Owned by the authors, published by EDP Sciences, 2013
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 2.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.