Etalonnage de transmetteur ou sonde portable H2O2 par l’utilisation des propriétés physico-chimiques de l’H2O2
Calibration of H2O2 transmitter or portable probe through the use of physical and chemical properties of H2O2
ARECO, 114 chemin de Saint Marc 06130 Grasse, France
VIRBAC, 13 ième rue LID BP27 06511 Carros Cedex, France
a f.guerrin@areco.fr
b robert.borel@virbac.com
c frederic.serre@virbac.com
Abstract
Decontamination of clean rooms and air treatment systems does not occur anymore in the pharmaceutical industry with formaldehyde, but with hydrogen peroxide or hydrogen peroxide plus adjuvants. Monitoring systems for biocide generation occurs through specific sensors tolerating these concentrated gases. This article describes a simple and reproducible method of calibration for this type of sensors, based on the calculation of the partial pressure of the H2O2 gas above a solution of H2O2 liquid for a known concentration at a known temperature. The developed method reduces the calibration uncertainty over current techniques used in industry.